Leica Arcadia H E105 Error — Cold Spot Not Cooling (“Peltier No Temp Reached” Fix)

Overview

If your Leica HistoCore Arcadia H displays an E105 Peltier cool: No temp reached (1h) error, the unit is unable to cool the cold spot area to its set temperature. This stops proper paraffin block cooling and stalls embedding workflow.

At Rankin Services, we frequently resolve this issue. One of the most common causes is a failed NTC temperature sensor in the Peltier cooling circuit.

When the sensor fails, the system never receives correct temperature feedback, so it cannot activate the fan or drive the Peltier module — resulting in a cold spot that’s warm and a timeout alarm.

Common Symptoms

  • E105 displayed on touchscreen
  • Cold spot stays warm or room temp
  • Cooling fan does not run
  • Peltier does not activate
  • Rest of system heats normally
  • Paraffin blocks fail to solidify

Most Likely Cause

Failed NTC temperature sensor inside the Peltier assembly

The Leica Arcadia H uses a Negative Temperature Coefficient (NTC) thermistor to monitor cold-spot temperature. Heat, age, and thermal cycling can cause resistance drift, open/failed thermistor, and incorrect temperature readings or no signal. When the NTC fails:

  • Controller receives no temperature feedback
  • Cooling fan will not start
  • Peltier module never drives cooling cycle
  • Unit times out after 1 hour and throws E105

How to Diagnose

Safety Note: Disconnect power before servicing internal assemblies.

  1. Power on and attempt cold-spot cooling cycle
  2. Verify cold spot does not drop in temperature
  3. Observe that fan remains off
  4. Confirm Peltier does not activate
  5. Inspect connections to Peltier module
  6. Measure NTC sensor reading for out-of-spec value or open circuit

If both the fan and Peltier are inactive and the cold spot stays warm — the NTC thermistor is the primary suspect.

Recommended Repair

Replace the NTC temperature sensor in the Peltier assembly

After NTC replacement:

  • Fan runs normally
  • Peltier cools cold spot
  • Temperature setpoint is reached
  • E105 error clears automatically

Pro Tip: Use the correct-value NTC. Using the wrong resistance spec may cause incorrect readings and recurring E105 faults.

Parts Required (OEM-Equivalent)

Peltier Temp Sensor 10488

Contact us for the correct thermistor specification and ordering support.

Repair Time & Skill Level

  • 60 – 120 minutes
  • Intermediate electronics troubleshooting

We provide depot service and onsite repair options.

Preventative Tip

As part of preventative maintenance:

  • Ensure cooling fan spins freely
  • Inspect connectors and cable harness
  • Verify rapid cooling response when cold-spot mode is activated

This helps catch sensor degradation early.